Unidirectional valve malfunction by the breakage or malposition of disc - two cases report -

机译:阀瓣破损或错位导致单向阀故障-两例报告-

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Malfunction of the unidirectional valve in a breathing circuit system may cause hypercapnia from the rebreathing of expired gas, ventilation failure, and barotrauma. Capnography is a useful method for monitoring the integrity of the unidirectional valve. We experienced two cases of malfunction of a unidirectional valve which caused leakage and reverse flow, diagnosed early as a change of the capnographic waveform. One case was caused by expiratory unidirectional valve breakage. The other was caused by an incorrectly-assembled inspiratory unidirectional valve.
机译:呼吸回路系统中的单向阀故障可能会由于呼出的气体重新呼吸,通气失败和气压伤而导致高碳酸血症。二氧化碳图是一种用于监控单向阀完整性的有用方法。我们经历了两例单向阀故障的情况,这些故障导致泄漏和逆流,早期诊断为二氧化碳图波形变化。一例是由于呼气单向瓣膜破裂引起的。另一个是由于吸气单向阀安装不正确引起的。