Usability of devices for self-injection: results of a formative study on a new disposable pen injector

机译:自我注射设备的可用性:新型一次性笔式注射器的形成性研究结果

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

This article presents a late-stage formative usability study of a pen-injector platform device. Such devices are used for the subcutaneous delivery of biopharmaceuticals, primarily for self-administration by the patient. The study was conducted with a broad user population, defined to represent user characteristics across a range of indications. The goals of the study were to confirm that the pen could be used without recurring patterns of use errors leading to hazardous situations, to evaluate the comprehension of the instructions for use (IFU), and to determine if training is necessary. In the study, a total of 36 participants in six groups (health care providers, caregivers, adolescents, diabetics with retinopathy, diabetics with neuropathy, and patients with arthritis) each read the IFU, prepared the device, and performed two simulated injections into an injection pad. Any use errors, near misses, or deviations from the IFU procedure were recorded. The overall success rate (injection completed by the participant without need for assistance) was 94% for the first and 100% for the second injection. Ninety-two percent of the participants reported that they felt confident using the device, 100% found the IFU helpful, and 75% found the device positively comfortable to use. Overall, a total average of 3.35 deviations and errors per user and injection were recorded (there were no near misses). Subtracting the errors without any potential for negative consequences for the injection or the user (trivial deviations), as well as those related to attaching and removing the pen needle (independent of the design of the pen itself), led to an average of 1.31 potentially relevant deviations per user and injection. It was concluded that the pen injector together with the IFU could be safely and efficiently used by all user groups without any training, and thus that the device and IFU in their current form are well suited for use in a range of specific applications.
机译:本文介绍了笔式注射器平台设备的后期格式化可用性研究。此类装置用于皮下递送生物药物,主要用于患者的自我给药。这项研究是针对广泛的用户群进行的,这些人群定义为代表各种适应症的用户特征。该研究的目的是确认可以使用笔而不会导致危险情况的重复使用错误模式,评估对使用说明(IFU)的理解,并确定是否需要培训。在该研究中,六个组(医疗保健提供者,护理人员,青少年,糖尿病性视网膜病变,糖尿病性神经病变和关节炎患者)的总共36位参与者各自阅读了IFU,准备了该设备,并进行了两次模拟注射。注射垫。记录任何使用错误,接近遗漏或与IFU程序的偏差。第一次注射的总体成功率(参与者完成注射而无需帮助)为94%,第二次注射为100%。 92%的参与者表示他们对使用该设备充满信心,100%认为IFU很有帮助,75%认为该设备使用起来非常舒适。总体上,每位用户和每次注射的平均偏差和错误总计为3.35(没有差错)。减去错误而不会给注射或使用者带来负面影响(微小偏差),以及与安装和拆卸笔式针头相关的错误(与笔本身的设计无关),则平均潜在值为1.31每个使用者和注射的相关偏差。结论是,笔式注射器和IFU可以在没有任何培训的情况下被所有用户组安全有效地使用,因此,当前形式的设备和IFU非常适合在各种特定应用中使用。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号