首页> 美国卫生研究院文献>IEEE Journal of Translational Engineering in Health and Medicine >A System for Simple Real-Time Anastomotic Failure Detection and Wireless Blood Flow Monitoring in the Lower Limbs
【2h】

A System for Simple Real-Time Anastomotic Failure Detection and Wireless Blood Flow Monitoring in the Lower Limbs

机译:下肢简单实时吻合术失败检测和无线血流监测系统

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Current totally implantable wireless blood flow monitors are large and cannot operate alongside nearby monitors. To alleviate the problems with the current monitors, we developed a system to monitor blood flow wirelessly, with a simple and easily interpretable real-time output. To the best of our knowledge, the implanted electronics are the smallest in reported literature, which reduces bio-burden. Calibration was performed across realistic physiological flow ranges using a syringe pump. The device’s sensors connected directly to the bilateral femoral veins of swine. For each 1 min, blood flow was monitored, then, an occlusion was introduced, and then, the occlusion was removed to resume flow. Each vein of four pigs was monitored four times, totaling 32 data collections. The implant measured 1.70 cm3 without battery/encapsulation. Across its calibrated range, including equipment tolerances, the relative error is less than ±5% above 8 mL/min and between −0.8% and +1.2% at its largest calibrated flow rate, which to the best of our knowledge is the lowest reported in the literature across the measured calibration range. The average standard deviation of the flow waveform amplitude was three times greater than that of no-flow. Establishing the relative amplitude for the flow and no-flow waveforms was found necessary, particularly for noise modulated Doppler signals. Its size and accuracy, compared with other microcontroller-equipped totally implantable monitors, make it a good candidate for future tether-free free flap monitoring studies.
机译:当前的完全可植入的无线血流监测器很大,不能与附近的监测器一起使用。为了减轻当前监视器的问题,我们开发了一种系统,可通过简单易懂的实时输出无线监控血液流动。据我们所知,植入的电子设备是文献报道中最小的电子设备,可减少生物负荷。使用注射泵在现实的生理流量范围内进行校准。该设备的传感器直接连接到猪的双侧股静脉。每隔1分钟监测一次血流,然后引入阻塞,然后去除阻塞以恢复流量。对四头猪的每条静脉进行了四次监测,总共收集了32个数据。植入物的尺寸为1.70 cm 3 ,没有电池/封装。在其校准范围内,包括设备公差,在最大校准流速下,相对误差在8 mL / min时小于±5%,在-0.8%至+ 1.2%之间,据我们所知,这是最低的报告值。在文献中测量的校准范围内。流量波形幅度的平均标准偏差比无流量大三倍。发现有必要建立流动和无流动波形的相对幅度,特别是对于噪声调制的多普勒信号。与其他配有微控制器的完全植入式监护仪相比,其尺寸和准确性使其成为未来无系链自由皮瓣监护研究的理想选择。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号