An audit of 2509 patient specimens analyzed for both total thyroxine (TT4) and free thyroxine (FT4) by the ACS-180 automated chemiluminescence immunoassay analyzer revealed that there were 219 discrepancies (8.7% of the total). A discrepancy was defined as one analyte within its reference range and the other outside. The discrepant results were divided into 4 groups: group A: normal TT4, but decreased FT4, 101 patients (4.0%); group B: increased TT4, normal FT4, 78 patients (3.1%); group C: decreased TT4, normal FT4, 34 patients (1.4%); and group D: normal TT4, increased FT4, 6 patients (0.2%). TSH measurements were available, by a 3rd generation chemiluminescent assay, in 142 of these patients, and were consistent with the FT4 result in 72 patients, with TT4 in 61 cases and with neither in 9 patients. The clinical diagnosis was investigated in a subgroup of 43 endocrine patients. Thirteen of the 20 endocrine patients in group A were diagnosed as hypothyroid, with a measured serum TSH, in 11 of them, of median 14.6, range 1.2 to 46.2 μlU/ml. Eleven of the 19 endocrine patients in group B were on thyroid replacement, with a measured serum TSH, in 7 of them of <0.01 μlU/ml. The audit of current laboratory practice led to a suggestion to replace the current thyroid function screening strategy of measuring both TT4 and FT4 by the combination of FT4 and TSH. The reasons for the discrepancies and the alternative strategies for screening of thyroid function are discussed.
机译:通过ACS-180自动化学发光免疫分析仪对2509名患者样本的总甲状腺素(TT4)和游离甲状腺素(FT4)进行了审计,结果发现有219个差异(占总数的8.7%)。差异定义为一种分析物在其参考范围内,另一种在外部。结果不一致分为4组:A组:TT4正常,但FT4降低,101例(4.0%); A组:TT4下降,FT4下降。 B组:TT4升高,FT4正常,78例(3.1%); C组:TT4下降,FT4正常,34例(1.4%); D组:TT4正常,FT4增加,6例(0.2%)。在其中的142例患者中,通过3代化学发光法可进行TSH测量,与72例患者的FT4结果一致,其中61例患者中的TT4结果与9例患者中均没有。在43名内分泌患者的亚组中调查了临床诊断。 A组20例内分泌患者中有13例被诊断为甲状腺功能减退,其中11例的血清TSH水平中位数为14.6,范围为1.2至46.2μL/ ml。 B组的19名内分泌患者中有11名接受了甲状腺置换术,并具有测得的血清TSH值,其中7名<0.01μL/ ml。对当前实验室实践的审核导致提出了通过结合FT4和TSH来代替同时测量TT4和FT4的甲状腺功能筛查策略的建议。讨论了差异的原因和筛查甲状腺功能的替代策略。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号