Comparison of Two Commercial Microimmunofluorescence Kits and an Enzyme Immunoassay Kit for Detection of Serum Immunoglobulin G Antibodies to Chlamydia pneumoniae

机译:两种商用微免疫荧光试剂盒和酶免疫分析试剂盒检测肺炎衣原体血清免疫球蛋白G抗体的比较

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

We compared the MRL and the Labsystems Chlamydia pneumoniae microimmunofluorescence (MIF) immunoglobulin G (IgG) kits and the Labsystems enzyme immunoassay (EIA) kit in a blinded study of 83 serum samples in which we evaluated titers, cross-reactivity to other species, and reproducibility. There was no statistically significant difference between the MRL and the Labsystems MIF kits in the endpoint titers of IgG antibody to C. pneumoniae. The correlation between the results obtained with these two MIF kits was excellent (r = 0.95; P = 0.001). The cross-reactivity of the C. pneumoniae-positive sera with C. trachomatis- and C. psittaci-positive sera was assessed for each MIF kit. For C. pneumoniae-positive sera with titers of ≥32, the Labsystems MIF kit exhibited more cross-reactivity to C. psittaci than the MRL kit did. The values obtained with the Labsystems EIA kit represented single dilutions of serum specimens expressed as enzymeimmuno units on a continuous scale. The results obtained with the Labsystems EIA kit correlated moderately well with those obtained with each MIF kit when they were compared for their abilities to detect IgG antibodies to C. pneumoniae (for the MRL MIF kit, r = 0.79 [P = 0.001]; for the Labsystems MIF kit, r = 0.78 [P = 0.001]). The results obtained with the commercial MRL and Labsystems MIF kits and the Labsystems EIA kit tested were reproducible; and the kits were standardized, had quality control reagents, and are suitable for detection of C. pneumoniae antibodies in serum and for use in interlaboratory studies. Validation of the use of these kits for clinical diagnosis still needs further evaluation.
机译:我们在83个血清样本的盲法研究中比较了MRL和Labsystems肺炎衣原体肺炎微免疫荧光(MIF)免疫球蛋白G(IgG)试剂盒和Labsystems酶免疫测定(EIA)试剂盒,在其中我们评估了效价,与其他物种的交叉反应性,以及重现性。在针对肺炎衣原体的IgG抗体的终点滴度中,MRL和Labsystems MIF试剂盒之间在统计学上无显着差异。用这两种MIF试剂盒获得的结果之间的相关性非常好(r = 0.95; P = 0.001)。对于每个MIF试剂盒,评估了肺炎衣原体阳性血清与沙眼衣原体和鹦鹉热衣原体阳性血清的交叉反应性。对于滴度≥32的肺炎衣原体阳性血清,Labsystems MIF试剂盒对鹦鹉热衣原体的交叉反应性比MRL试剂盒大。用Labsystems EIA试剂盒获得的值代表连续稀释的以酶免疫单位表示的血清样品的单倍稀释度。将Labsystems EIA试剂盒获得的结果与每种MIF试剂盒获得的结果进行比较,比较它们检测肺炎衣原体IgG抗体的能力(对于MRL MIF试剂盒,r = 0.79 [P = 0.001];对于Labsystems MIF套件,r = 0.78 [P = 0.001])。使用商业MRL和Labsystems MIF套件以及经测试的Labsystems EIA套件所获得的结果是可重复的;试剂盒经过标准化处理,具有质量控制试剂,适合检测血清中的肺炎衣原体抗体,并用于实验室间研究。这些试剂盒用于临床诊断的验证仍需进一步评估。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号