Comparison of performance of various sphygmomanometers with intra-arterial blood-pressure readings.

机译:各种血压计的性能与动脉内血压读数的比较。

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Seven types of sphygmomanometer were used in random order on each of nine hypertensive patients and the readings compared with simultaneous intra-arterial blood-pressure recordings. All the devices gave significantly different values for systolic pressure, and only two measured diastolic pressure without significant error. Systolic pressure was consistently underestimated (range 31-7 mm Hg), and all but one instrument overestimated diastolic pressure (range 10-2 mm Hg). The variability of readings was least with the standard mercury sphygmomanometer and the random-zero machine, while with some of the more automated devices single readings were in error up to -68/33 mm Hg. The strong correlations found between intra-arterial and cuff systolic pressures with all devices tested and significant correlations for diastolic pressure with all but one device indicate that, with one possible exception, the sphygmomanometers would give accurate results where a change in blood pressure was the main concern.
机译:在9名高血压患者中,每种患者随机使用7种血压计,并将其读数与同时进行的动脉内血压记录进行比较。所有设备的收缩压值都明显不同,只有两个测得的舒张压没有明显误差。收缩压始终被低估(范围为31-7 mm Hg),除一台仪器外,所有器械均高估了舒张压(范围为10-2 mm Hg)。标准水银血压计和随机零仪的读数变化最小,而某些自动化程度更高的设备的单个读数的误差高达-68/33 mm Hg。在所有测试的设备中,动脉内和袖带收缩压之间的强相关性;除一种设备外,与所有设备的舒张压之间的显着相关性表明,除了血压可能是主要因素外,血压计会给出准确的结果关心。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号