首页> 美国卫生研究院文献>Biochemia Medica >Elevated zinc concentrations in a 5 months old infant: A case report
【2h】

Elevated zinc concentrations in a 5 months old infant: A case report

机译:5个月大婴儿锌含量升高:一例报告

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Pre-analytical errors account for the majority of laboratory-associated errors. In a 5 months old infant hospitalised with lung dysfunction due to prematurity, a routine measurement of zinc revealed an unexpected elevated concentration of 20.2 µmol/L (reference interval 10.0 - 19.0 µmol/L) compared to 11.6 µmol/L five days earlier. Zinc measurement was repeated two days later and had further increased to 42.4 µmol/L. Of note, there were no clinical signs of the increased zinc concentrations. Performance data for the zinc analysis (performed by inductively coupled plasma mass spectrometry) was found satisfactory. A thorough review of the patient´s medication and nutrition supplements revealed no relevant zinc content. The blood was obtained through capillary blood sampling, and anything at the skin puncture site containing zinc could therefore potentially contribute to the elevated zinc results. It was investigated if any ointment containing zinc had been applied at the puncture site, which revealed that the mother had applied vitamin E ointment containing zinc-oxide at the infant’s heel. A capillary sample obtained from the opposite heel, where no vitamin E ointment had been applied, revealed a zinc concentration of 14.3 µmol/L. In conclusion, pre-analytical contamination with ointments must be considered in case of unexpected measurements from capillary blood.
机译:分析前错误占大多数与实验室相关的错误。在一个因早产而导致肺功能不佳住院的5个月大婴儿中,锌的常规检测显示出意料之外的20.2 µmol / L浓度升高(参考区间10.0-19.0 µmol / L),而五天前为11.6 µmol / L。两天后重复进行锌测量,并进一步提高到42.4 µmol / L。值得注意的是,没有锌浓度升高的临床迹象。发现锌分析的性能数据(通过电感耦合等离子体质谱法执行)令人满意。对患者的药物和营养补品进行了彻底的检查后发现,锌含量不相关。血液是通过毛细血管采血获得的,因此皮肤穿刺部位的任何含锌物质都可能导致锌含量升高。对穿刺部位是否使用了含锌软膏进行了调查,结果表明母亲在婴儿脚后跟使用了含氧化锌的维生素E软膏。从未施加维生素E软膏的对侧脚跟获得的毛细管样品显示锌浓度为14.3 µmol / L。总之,如果从毛细血管血中进行意外测量,则必须考虑使用软膏进行分析前污染。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号