首页> 外文期刊>等离子体科学和技术(英文版) >Dynamic Performance of the ITER Reactive Power Compensation System
【24h】

Dynamic Performance of the ITER Reactive Power Compensation System

机译:ITER无功补偿系统的动态性能ITER无功补偿系统的动态性能

获取原文
获取原文并翻译 | 示例
       

摘要

Dynamic performance of a reactive power compensation (RPC) system for the international thermonuclear experimental reactor (ITER) power supply is presented. Static var compensators (SVCs) are adopted to mitigate voltage fluctuation and reduce the reactive power down to a level acceptable for the French/European 400 kV grid. A voltage feedback and load power feedforward controller for SVC is proposed, with the feedforward loop intended to guarantee short response time and the feedback loop ensuring good dynamics and steady characteristics of SVC. A mean filter was chosen to measure the control signals to improve the dynamic response. The dynamic performance of the SVC is verified by simulations using PSCAD/EMTDC codes.
机译:介绍了国际热核实验堆(ITER)电源的无功补偿(RPC)系统的动态性能。采用静态无功补偿器(SVC)来减轻电压波动并将无功功率降低到法国/欧洲400 kV电网可接受的水平。提出了一种用于SVC的电压反馈和负载功率前馈控制器,其前馈环路旨在确保较短的响应时间,而反馈环路则确保了SVC的良好动态特性和稳定特性。选择一个平均滤波器来测量控制信号以改善动态响应。通过使用PSCAD / EMTDC代码进行仿真,可以验证SVC的动态性能。

著录项

获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号