首页> 中文期刊> 《核技术:英文版》 >Determination of water equivalent ratio for some dosimetric materials in proton therapy using MNCPX simulation tool

Determination of water equivalent ratio for some dosimetric materials in proton therapy using MNCPX simulation tool

         

摘要

The water equivalent ratio(WER) was calculated for polypropylene(PP), paraffin, polyethylene(PE), polystyrene(PS), polymethyl methacrylate(PMMA), and polycarbonate materials with potential applications in dosimetry and medical physics. This was performed using the Monte Carlo simulation code, MCNPX, at different proton energies. The calculated WER values were compared with National Institute of Standards and Technology(NIST) data, available experimental and analytical results,as well as the FLUKA, SRIM, and SEICS codes. PP and PMMA were associated with the minimum and maximum WER values, respectively. Good agreement was observed between the MCNPX and NIST data. The biggest difference was 0.71% for PS at 150 MeV proton energy. In addition, a relatively large positive correlation between the WER values and the electron density of the dosimetric materials was observed. Finally, it was noted that PE presented the most analogous Depth Dose Characteristics to liquid water.

著录项

获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号