首页> 中文期刊> 《核技术:英文版》 >Two-phase flow component fraction measurement using gamma-ray attenuation technique

Two-phase flow component fraction measurement using gamma-ray attenuation technique

         

摘要

Multiphase flow meters as the potential alternatives to separation and metering techniques have been in rapid development since 1980 s.Before its field operation,the instrument should be calibrated in a standard test-facility.In spite of the known medium and large scale facilities all over the world,we developed a laboratory scale instrument for component fraction measurements.It has a two-phase flow homogenizer loop with the clamp-on potential of the meters to provide a regime independent measurement.It is capable of delivering a complete homogenization by γ-ray densitometer.With an error of±5%in component fraction measurements,this instrument is appropriate for testing and calibrating other meters.

著录项

获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有