首页> 中文期刊> 《核技术:英文版》 >Alignment of beam position monitors in cryomodule of CADS injector Ⅱ

Alignment of beam position monitors in cryomodule of CADS injector Ⅱ

         

摘要

Significant temperature difference(300-77 K or even 4 K) can cause large deformations and displacements of the beam position monitors(BPMs),which affect BPMs measurement resolution or even cause their malfunction in cryogenic situations.In this paper,to check the offset from the mechanical to electrical center in low temperature(77 K),Fourier's law and finite element method are used to simulate cryo-deformation.Laser tracker and micro-alignment telescope are employed in combined BPM calibration,installation and monitoring.The calibration error is<0.02 mm,and the installation and monitoring precision are 0.06 mm and 0.01 mm,respectively.The monitored cryo-deformation agrees well with the simulation results.These indicate that the combined alignment can improve performance of the BPM system.All these guaranteed the success of running the 9.55 MeV@2.14 mA cw protons.

著录项

获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有