首页> 中文期刊>中国环境监测 >固体进样-冷原子吸收法直接测定土壤中总汞

固体进样-冷原子吸收法直接测定土壤中总汞

     

摘要

应用固体测汞仪,采用多标准土壤样品( n>20)绘制校准曲线法和单一标准土壤样品绘制校准曲线法,分别对土壤中的汞进行测定。结果表明,两者的方法检出限分别为0.30、1.49 ng,平行样(n=6)的相对标准偏差分别为3.6%~4%、5.4%~9.0%,90 d内的重复性精密度为4.1%,对国家土壤标样进行测定,结果与标准值相符。表明与单一标准土壤样品绘制校准曲线法相比,多标准土壤样品绘制标准曲线法具有更好的精密度、更低的检出限,更强的适用性,并且由于校准曲线长期稳定性,有效缩短土壤中汞的检测周期。%Total mercury in the state standard soil samples were determined using direct analyzer with solid sampling by multi-state standard soil-calibration curve and single-state standard soil-calibration curve. The results showed that the detection limits were 0. 30 and 1. 49 ng, the relative standard deviations of parallel samples ( n =6 ) were 3. 6%-4% and 5. 4%-9. 0%, the repeatability precisions in 90 days was 4. 1%. This suggested that the method of multi-state standard soil-calibration curve had the lower detection limit,the better replicability precision,and the better reproducibility precision.

著录项

  • 来源
    《中国环境监测》|2016年第3期|126-128|共3页
  • 作者单位

    河南省环境监测中心;

    河南 郑州 450004;

    河南省环境监测中心;

    河南 郑州 450004;

    河南省环境监测中心;

    河南 郑州 450004;

    河南省环境监测中心;

    河南 郑州 450004;

    河南省环境监测中心;

    河南 郑州 450004;

    河南省环境监测中心;

    河南 郑州 450004;

    河南省环境监测中心;

    河南 郑州 450004;

  • 原文格式 PDF
  • 正文语种 chi
  • 中图分类 监测分析方法;
  • 关键词

    汞; 固体测汞仪; 土壤;

  • 入库时间 2022-08-17 18:11:10