首页> 中文期刊>中国医学装备 >高频电刀的质量控制方法与检测结果分析

高频电刀的质量控制方法与检测结果分析

     

摘要

目的:根据高频电刀检测标准建立检测方法,测定高频电刀的电气安全等数据,进行成因分析,制订解决方案.方法:使用FLUKE电气安全测试仪和高频电刀测试仪,对2015-2016年医院在用高频电刀进行电气安全、输出功率等性能指标检测,统计检测结果数据进行分析.结果:2015年和2016年高频电刀的质量控制检测合格率分别为79%和81%,对不合格的高频电刀做相应工程维护方案,消除其临床使用中存在的风险隐患.结论:建立的高频电刀检测方法有效用于质量控制检测.检测数据提示,要定期对高频电刀进行性能检测是重要的技术故障措施,以便掌握机器的性能状态,对存在风险的机器可及时发现和处理,避免安全隐患,增强对高频电刀的风险管理,使其安全有效地应用于临床手术.%Objective:To establish the quality control method of electrosurgical unit(ESU) through the safety performance test and data analysis for electrotome.Methods: According to the testing standard of ESU, Fluke ESA612 electrical safety analyzer and QA-ES II electrosurgery unit analyzer were applied to test the properties of ESU were using during 2015 and 2016, and the testing results were analyzed.Results:In 2015 and 2016, the qualified rate of quality control results were 79% and 81%, respectively; the problems of unqualified ESU were resolved and their hidden risks in clinical applications were eliminated.Conclusion: By testing the properties of ESU, we can master the performance of the machine; find and deal with the risk of machine in time, and avoid the hidden danger and enhance the risk management for ESU, so that the machine can be safely and effectively applied in clinical surgery.