[Objectives]A method for the detection of monensin in poultry and livestock meat by pre-column derivatization-high performance liquid chromatography was established.[Methods]The sample was extracted with chloroform,derivatized with trichloroacetic acid and 2,4-dinitrophenylhydrazine,and centrifuged to obtain a purified solution.A C18 chromatographic column(4.6 mm×150 mm,5μm)was used for separation with(1.5%)acetic acid water∶methanol(volume ratio)=1∶9 as the mobile phase using a DAD detector for detection,and the external standard method was adopted for peak area quantification.[Results]Monensin had good linearity in the concentration range of 5.00-200 mg/L,with the linear correlation coefficient r 2>0.999;the detection limit was 5.00 mg/kg;the relative standard deviation was smaller than 10%;and the recoveries of standard addition experiment were in the range of 75%-110%.[Conclusions]The method has the advantages of simple pretreatment operation,good derivatization effect and fast detection speed,and is suitable for detecting monensin in poultry and livestock meat.