Differences of the Mechanical Setup of Hip Simulators and Their Consequences on the Outcome of Hip Wear Testing

机译:髋部模拟器机械设置的差异及其对髋部磨损测试结果的影响

获取原文
获取原文并翻译 | 示例

摘要

The search for improved wear resistant materials for hip endoprostheses has lead to an international standard for wear testing of total hip joint prosthesis (ISO 14242). The displacement and load curves in this standard are derived from clinical gait analysis. However the standard leaves the interpretation of its displacement curves open since no coordinate system is defined in the document. By reviewing the development of the standard, the apparent coordinate system is identified and its application from both, an anatomical point of view and a technical perspective for designing hip simulators is explained. The study examines the coordinate transformations necessary for using common gait analysis data with the standard. On the basis of these transformations, the relative movements of the articulating surfaces are compared with existing hip simulators, the situation in vivo and the guidelines of the ISO standard. Thus possible consequences on the outcome of hip wear testing are discussed.
机译:寻找用于髋关节假体的改进的耐磨材料已经导致了整个髋关节假体的磨损测试的国际标准(ISO 14242)。本标准中的位移和载荷曲线来自临床步态分析。但是,由于在文档中未定义坐标系,因此该标准对其位移曲线的解释仍然是开放的。通过回顾该标准的制定,可以识别视在坐标系,并从解剖学的角度和设计髋关节模拟器的技术角度对视坐标系进行了说明。该研究检查了使用标准步态分析数据进行必要的坐标转换。在这些转换的基础上,将关节表面的相对运动与现有的髋关节模拟器,体内情况和ISO标准指南进行了比较。因此,讨论了髋关节磨损测试结果的可能后果。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号