The precision of AAPM TG-43 dose calculation formalism in heterogeneous media has been studied based on the survey and evaluation of the status of interstitial brachytherapy. After a brief introduction of radioactive seeds (125I, 192Ir and 103Pd) and AAPM TG-43 formalism, the radial dose rates calculated by Monte Carlo code MCNP-4C and TG-43 formalism were compared with homogeneous and heterogeneous phantoms. The reason of errors between the results of the above two methods was analyzed, and modified dose rates constant were proposed and testified. With the updated dose rates constant, AAPM TG-43 formalism can predict the dose rate distribution accurately and rapidly in clinical.
机译:基于对间质近距离放射治疗现状的调查和评估,研究了AAPM TG-43剂量计算形式在不同介质中的精度。在简要介绍放射性种子(125I,192Ir和103Pd)和AAPM TG-43形式主义之后,将蒙特卡罗代码MCNP-4C和TG-43形式主义计算出的径向剂量率与同质和异质体模进行了比较。分析了上述两种方法的结果之间存在误差的原因,并提出并证明了修正的剂量率常数。在更新的剂量率恒定的情况下,AAPM TG-43形式可以在临床中准确,快速地预测剂量率分布。