Glucose measurements in solutions using fiber optic evanescent wave spectroscopy and tunable CO2 laser

机译:使用光纤van逝波光谱和可调谐CO2激光在溶液中进行葡萄糖测量

获取原文
获取原文并翻译 | 示例

摘要

Abstract: There is an interest in using IR spectroscopy techniques for blood analysis, and in particular for glucose measurements. Glucose concentration in aqueous solutions were measured using fiberoptic evanescent wave spectroscopy. A new spectrometer consisting of tunable CO$-2$/ laser as the radiation source, and a silver halide fiber was built. Several solutions were measured for calibration with concentrations in the range of 70-2000 mg/dl. After the calibration, solution were measured and predicted according to the calibration parameters. The predicted concentrations were in agreement with the reference ones. The predicted errors for concentrations range up to 200 mg/dl, which is the clinical range, was less than 18 mg/dl. The AgClBr fibers are non-toxic and therefore this measurement can be carried out in real-time, for in-situ measurements of glucose in blood. !10
机译:摘要:人们对使用红外光谱技术进行血液分析,特别是对葡萄糖测量感兴趣。使用光纤e逝波光谱法测量水溶液中的葡萄糖浓度。建造了一种新的光谱仪,该光谱仪由可调的CO $ -2 $ /激光作为辐射源,以及卤化银纤维组成。测量了几种溶液用于校准,浓度范围为70-2000 mg / dl。校准后,根据校准参数测量溶液并进行预测。预测浓度与参考浓度一致。浓度范围的预测误差最高为200 mg / dl,这是临床范围,小于18 mg / dl。 AgClBr纤维是无毒的,因此可以实时进行此测量,以进行血液中葡萄糖的原位测量。 !10