首页> 外文会议>Society of Photo-Optical Instrumentation Engineers (SPIE);SPIE Proceedings >Characterization of Scatter Radiation of a Breast Phantom on Siemens Prototype FFDM with and without an Anti-Scatter Grid
【24h】

Characterization of Scatter Radiation of a Breast Phantom on Siemens Prototype FFDM with and without an Anti-Scatter Grid

机译:带有和不带有防散射网格的西门子原型FFDM上乳房幻影的散射辐射特征

获取原文

摘要

In this study, the beam stop technique was applied to obtain the scatter fraction values for an anthropomorphic breastphantom on a flat-panel full field mammography system. The phantom was equivalent to a compressed breast of 5 cmthickness with 50% glandular tissue content. The images were acquired at 28kVp with Mo/Mo target/filter combinationand multiple mAs values with or without an anti-scatter grid. The one-dimensional and two-dimensional scatter fractionprofiles of the breast phantom without a grid were plotted. The effect of mAs on scatter fraction calculation wasinvestigated. It was found that in order to get a reliable measurement of scatter fraction, the mAs had to be above about40mAs without a grid and 160mAs with a grid. In addition, the SNR values at 28kVp and 80mAs (AEC level for thephantom using the same imaging technique on a screen/film system) with and without a grid were compared with eachother. The SNR with a grid was slightly smaller than that without a grid. The SNR improvement factor (KSNR) defined asthe ratio of SNR without the grid to SNR with the grid was 0.976. The grid had the primary and scatter transmissions of73% and 13% respectively.
机译:在这项研究中,使用束停止技术来获得平板全视野乳房X线照相系统上拟人化乳房假体的散射分数值。体模相当于5厘米厚,50%腺体组织含量的受压乳房。图像是在28kVp下使用Mo / Mo目标/滤光片组合以及带有或不带有防散射网格的多个mAs值采集的。绘制没有网格的乳房体模的一维和二维散射分数分布图。研究了mAs对散射分数计算的影响。已经发现,为了获得对散射分数的可靠测量,不带电网的mAs必须高于约40mAs,而带电网的mAs必须高于160mAs。此外,将具有和不具有网格的28kVp和80mAs(在屏幕/胶片系统上使用相同成像技术的幻影的AEC电平)的SNR值进行了相互比较。有网格的信噪比略小于无网格的信噪比。定义为不带网格的SNR与带网格的SNR之比的SNR改善因子(KSNR)为0.976。网格的一次传输和分散传输分别为73%和13%。

著录项

  • 来源
  • 会议地点
  • 作者单位

    Department of Biomedical Engineering Duke University Durham NC USA 27708 Duke Advanced Imaging Laboratories Department of Radiology Duke University Medical Center Durham NC USA 27710 Institute of Statistics and Decision Sciences Duke University Durham NC USA 27708;

    Department of Biomedical Engineering Duke University Durham NC USA 27708 Duke Advanced Imaging Laboratories Department of Radiology Duke University Medical Center Durham NC USA 27710;