首页> 外文会议>Single molecule spectroscopy and imaging IV >Ultra high-throughput single molecule spectroscopy with a 1024 pixel SPAD
【24h】

Ultra high-throughput single molecule spectroscopy with a 1024 pixel SPAD

机译:具有1024像素SPAD的超高通量单分子光谱

获取原文
获取原文并翻译 | 示例

摘要

Single-molecule spectroscopy is a powerful approach to measuring molecular properties such as size, brightness, conformation, and binding constants. Due to the low concentrations in the single-molecule regime, measurements with good statistical accuracy require long acquisition times. Previously we showed a factor of 8 improvement in acquisition speed using a custom-CMOS 8x1 SPAD array. Here we present preliminary results with a 64X improvement in throughput obtained using a liquid crystal on silicon spatial light modulator (LCOS-SLM) and a novel standard CMOS 1024 pixel SPAD array, opening the way to truly high-throughput single-molecule spectroscopy.
机译:单分子光谱法是一种测量分子性质的有效方法,例如大小,亮度,构象和结合​​常数。由于单分子浓度低,具有良好统计准确性的测量需要较长的采集时间。以前,我们使用定制CMOS 8x1 SPAD阵列显示出采集速度提高了8倍。在这里,我们展示了使用硅空间液晶调制器(LCOS-SLM)和新型标准CMOS 1024像素SPAD阵列获得的吞吐量提高64倍的初步结果,为真正的高通量单分子光谱学开辟了道路。