首页> 外文会议>Sensors, Systems, and Next-Generation Satellites VIII >Calibration Methodology for the Airborne Dispersive Pushbroom Imaging Spectrometer (APEX)
【24h】

Calibration Methodology for the Airborne Dispersive Pushbroom Imaging Spectrometer (APEX)

机译:机载色散推扫成像光谱仪(APEX)的校准方法

获取原文
获取原文并翻译 | 示例

摘要

APEX is a dispersive pushbroom imaging spectrometer operating in the spectral range between 380 - 2500 nm. The spectral resolution will be better than 10 nm in the SWIR and < 5 nm in the VNIR range of the solar reflected range of the spectrum. The total FOV will be ± 14 deg, recording 1000 pixels across track with about 300 spectral bands simultaneously. A large variety of characterization measurements will be performed in the scope of the APEX project, e.g., on-board characterization, frequent laboratory characterization, and vicarious calibration. The retrieved calibration parameters will allow a data calibration in the APEX Processing and Archiving Facility (PAF). The data calibration includes the calculation of the required, time-dependent calibration coefficients from the calibration parameters and, subsequently, the radiometric, spectral and geometric calibration of the raw data. Because of the heterogeneity of the characterization measurements, the optimal calibration for each data set is achieved using a special assimilation algorithm. In the paper the different facilities allowing characterization measurements, the PAF and the new data assimilation scheme are outlined.
机译:APEX是一种色散的推扫式成像光谱仪,在380-2500 nm的光谱范围内运行。在SWIR中,光谱分辨率将优于10 nm,在光谱的太阳反射范围中,VNIR范围将小于5 nm。总FOV将为±14度,同时记录大约300个光谱带的1000个像素。在APEX项目的范围内将执行各种各样的表征测量,例如,机载表征,频繁的实验室表征和替代校准。检索到的校准参数将允许在APEX处理和存档设施(PAF)中进行数据校准。数据校准包括根据校准参数计算所需的,与时间有关的校准系数,然后进行原始数据的辐射,光谱和几何校准。由于表征测量的异质性,使用特殊的同化算法可实现每个数据集的最佳校准。本文概述了允许进行表征测量的不同设备,PAF和新的数据同化方案。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号