We demonstrated that a synthesized light source (SLS) consisting of two low coherence light sources is capable of enhancing spatial resolution in optical coherence tomography (OCT). The axial resolution of OCT is given by the half of coherence length of the light source. We fabricated an SLS with a coherence length of 2.3 μm and the side lobe intensity of 29 % under the intensity ratio of LED 1: LED 2 = 1: 0.4 by combining two light sources, LED 1 with a central wavelength of 700 nm and spectral bandwidth of 100 nm, and LED 2 of 898 run and 85 nm. The coherence length of 2.3 μm is 56% of the shorter coherence length in two LED's, therefore the axial resolution is 1.2 μm.
机译:我们证明了由两个低相干光源组成的合成光源(SLS)能够增强光学相干断层扫描(OCT)中的空间分辨率。 OCT的轴向分辨率由光源相干长度的一半给出。通过结合两个光源,即中心波长为700 nm的LED 1和光谱,在LED 1:LED 2 = 1:0.4的强度比下,我们制造出了相干长度为2.3μm,旁瓣强度为29%的SLS。带宽为100 nm,LED 2的运行频率为898 nm,波长为85 nm。 2.3μm的相干长度是两个LED中较短的相干长度的56%,因​​此轴向分辨率为1.2μm。