首页> 外文会议>Rare metals >Structural design and properties of the aorta stent-graft
【24h】

Structural design and properties of the aorta stent-graft

机译:主动脉覆膜支架的结构设计和性能

获取原文

摘要

A new shape memory alloy thoracic aorta stent-graft with taper,no girder structure designed for endovascular repair procedure was introduced in this paper. The radial support force of the stent was measured by electron universal test machine,the recovery tempera ture was measured by DSC,and structural design of the stent was verified by doctors. The results showed that the stent recovered com pletely at 25.9? and radial holding force of the stent distributed between 6 N and 8 N.Doctors operated more easily,and the stent was of better flexibility and uniform expandability,met the physiology characteristics of people's thoracic aorta more. The stent-graft with ta pered design may be fit for treatment option for endovascular repair procedure especially,and clinical treatment effects were satisfactory.
机译:本文介绍了一种新型的形状记忆合金胸主动脉覆膜支架,该支架具有锥形无梁结构,旨在用于血管内修复手术。用电子万能试验机测量支架的径向支撑力,用DSC测量恢复温度,并通过医生验证支架的结构设计。结果显示支架完全恢复到25.9?。支架的径向支撑力分布在6 N和8 N之间。医生操作起来更容易,并且支架具有更好的柔韧性和均匀的可扩展性,更符合人们胸主动脉的生理特性。锥形设计的覆膜支架尤其适合血管内修复手术的治疗选择,临床疗效满意。

著录项

  • 来源
    《Rare metals》|2009年|p.525-528|共4页
  • 会议地点 Jiaxing(CN)
  • 作者单位

    General Research Institutes for Non-Ferrous Metals,Grikin Advanced Materials Co.,Ltd.,Beijingl02200,China;

    General Research Institutes for Non-Ferrous Metals,Grikin Advanced Materials Co.,Ltd.,Beijingl02200,China;

    General Research Institutes for Non-Ferrous Metals,Grikin Advanced Materials Co.,Ltd.,Beijingl02200,China;

    Beijing Anzhen Hospital of the Capital University of Medical Sciences,Beijing 100029,China;

  • 会议组织
  • 原文格式 PDF
  • 正文语种
  • 中图分类 稀有金属和少量金属;
  • 关键词

    TiNi alloy; stent-graft; thoracic aorta; endovascular repair procedure;

    机译:TiNi合金;支架移植物;胸主动脉;血管内修复程序;