首页> 外文会议>Rare Metal Materials and Engineering vol.35 Suppl.1 February 2006: Technology and Applications of Chemical Sensors >Application of Dynamic Linear Detecting Method in Data Processing of the Portable Blood Sugar Analyzer
【24h】

Application of Dynamic Linear Detecting Method in Data Processing of the Portable Blood Sugar Analyzer

机译:动态线性检测方法在便携式血糖分析仪数据处理中的应用

获取原文
获取原文并翻译 | 示例

摘要

In this paper, a dynamic linear detecting method, that the non-linear coefficient NL% was led and the non-linearity of data were estimated continuously and dynamically and determined when NL% exceeded reference value (5%), was used for data processing and could solve the problem caused by the phenomenon of substrate depleting occurred following the redox reaction in portable blood sugar analyzer. By contrast to the conventional end-point method, the dynamic linear detecting method is based on multipoint data collecting. Experiments of measuring the calibration glucose solution with 8 various concentrations from 50 mg/dl to 400 mg/dl were carried out with the analyzer developed by our group. The linear regression curve, whose correlation for the data was 0.9995 and the residual was 2.8080, were obtained. The obtained correlation, residual, and the computation workload are all fit for the portable blood sugar analyzer.
机译:本文采用动态线性检测方法进行数据处理,该方法检测非线性系数NL%,并连续动态地估计数据的非线性,并确定NL%超过参考值(5%)时的非线性。解决了便携式血糖分析仪中的氧化还原反应后发生底物耗尽现象的问题。与传统的端点方法相比,动态线性检测方法基于多点数据收集。我们小组开发的分析仪进行了从50 mg / dl至400 mg / dl的8种不同浓度的校准葡萄糖溶液的测量实验。获得了线性回归曲线,其与数据的相关性为0.9995,残差为2.8080。所获得的相关性,残差和计算量都适合于便携式血糖分析仪。