首页> 外文会议>East Asian Chonference on Chemical Sensors >Application of Dynamic Linear Detecting Method in Data Processing of the Portable Blood Sugar Analyzer
【24h】

Application of Dynamic Linear Detecting Method in Data Processing of the Portable Blood Sugar Analyzer

机译:动态线性检测方法在便携式血糖分析仪数据处理中的应用

获取原文

摘要

In this paper, a dynamic linear detecting method, that the non-linear coefficient NL% was led and the non-linearity of data were estimated continuously and dynamically and determined when NL% exceeded reference value (5%), was used for data processing and could solve the problem caused by the phenomenon of substrate depleting occurred following the redox reaction in portable blood sugar analyzer. By contrast to the conventional end-point method, the dynamic linear detecting method is based on multipoint data collecting. Experiments of measuring the calibration glucose solution with 8 various concentrations from 50 mg/dl to 400 mg/dl were carried out with the analyzer developed by our group. The linear regression curve, whose correlation for the data was 0.9995 and the residual was 2.8080, were obtained. The obtained correlation, residual, and the computation workload are all fit for the portable blood sugar analyzer.
机译:在本文中,一种动态线性检测方法,即非线性系数N1%LED,并且当N1%超过参考值(5%)时,连续和动态地估计数据的非线性,并确定数据处理并且可以解决在便携式血糖分析仪中氧化还原反应后脱脂现象引起的问题。相反,与传统的终点方法相比,动态线性检测方法基于多点数据收集。用本组开发的分析仪进行测量含有80mg / d1至400mg / dl的校准葡萄糖溶液的实验。线性回归曲线,其对数据的相关性为0.9995,并且获得残留为2.8080。所获得的相关性,残差和计算工作量都适合便携式血糖分析仪。