首页> 外文会议>Rare Metal Materials and Engineering vol.35 Suppl.1 February 2006: Technology and Applications of Chemical Sensors >High Sensitivity Biomolecular Recognition Device Based on Giant Magnetoresistance Effect
【24h】

High Sensitivity Biomolecular Recognition Device Based on Giant Magnetoresistance Effect

机译:基于巨磁阻效应的高灵敏度生物分子识别装置

获取原文
获取原文并翻译 | 示例

摘要

The structure and microfabrication, the detecting theory and the way of biomolecular recognition device based on giant magnetoresistance(GMR) effect are introduced, also the signal detecting and processing instrumentation are presented. Here the GMR biosensor was fabricated with magnetic tunnel junction(MJT) material. The biomolecular recognition device contains an array of MJT sensors, single MJT sensor size is 10 μm × 20 μm, tunneling magnetoresistance ratio(TMR) at room temperature is 52.2%, the typical values of junction resistance-area product R_S is 2.6 kΩμm~2, detecting sensitivity of this system is about 8 × 10~(-4) A·m~(-1). Bioadaptation layer of this device was fabricated with PDMS the thickness of which is less than 100 nm.
机译:介绍了基于巨磁阻效应的生物分子识别装置的结构和微加工,检测原理和方法,并介绍了信号检测与处理仪器。在这里,GMR生物传感器是用磁隧道结(MJT)材料制造的。该生物分子识别装置包含MJT传感器阵列,单个MJT传感器尺寸为10μm×20μm,室温下的隧道磁阻比(TMR)为52.2%,结电阻面积乘积R_S的典型值为2.6kΩμm〜2 ,该系统的检测灵敏度约为8×10〜(-4)A·m〜(-1)。用PDMS制造该装置的生物适应层,其厚度小于100nm。