ANALYSIS OF PERFORMANCE PARAMETERS OF ILLUMINANCE METERS PER CIE DS 023 QUALITY INDICES FOR SPECIFIC FIELD MEASUREMENTS

机译:特定场测量中每个CIE DS 023质量指数的照度性能参数分析

获取原文
获取原文并翻译 | 示例

摘要

Illuminance meters and luminance meters are widely used instruments for many types of field measurements. Characterization of performance of these devices was formerly made according to CIE 69:1987 recently updated by the draft standard CIE DS 023/E:2012 which if would be approved by the CIE national commities, it will be published as a CIE Standard and later on as a joint ISO/CIE standard. But the problem of user as describe this draft standard is the characteristics of these photometers alone do not allow the estimation of the measurement uncertainty for a specific measurement task even if on the other hand it describes that instruments with "better" characteristics in most cases allow smaller uncertainties than instruments with "worse" properties. Often users do not know according to quality indices evaluate measurement uncertainties of their measurement especially in the field photometric measurements of illuminance or luminance. This ambiguity appears in the at the verification measurements of lighting systems after realisation of lighting project or others specific measurements in the field. This paper concerns about analysis of measurement uncertainties for specific field measurements of illuminance with different classes of illuminance meters defined by the draft standard CIE DS 023/E:2012. The main aim of research work was made about field measurements for indoor workplaces under well-known conditions. Lighting system parameters according to EN 12464-1 indoor work places (i.e. average maintained illuminance level E_m, uniformity U_0 etc.) and spectrum of used light sources (in luminaires) in the measured rooms was known to compare results of measurements with stated expanded uncertainties for 95% confidance interval (k=2) assuming of normal distribution. Uncertainties of measurement in this specific field measurement was investigated for different types of illuminance meters from worse classes to the best classes of quality assuming quality indices prescribed in the draft standard. From results of this work can be in some level to be helpful to predict the accuracy of measurement of used illuminance meters with concrete class to avoid ambiguity for user who performs field measurements with them.
机译:照度计和照度计是用于许多类型的现场测量的广泛使用的仪器。这些设备的性能表征以前是根据最近由标准草案CIE DS 023 / E:2012更新的CIE 69:1987进行的,如果该草案将得到CIE国家承诺的批准,则将作为CIE标准发布,随后发布。作为ISO / CIE联合标准。但是,描述该标准草案的用户所面临的问题是,仅这些光度计的特性就无法估算特定测量任务的测量不确定度,即使另一方面,它描述了在大多数情况下具有“更好”特性的仪器比具有“更差”属性的工具具有更小的不确定性。通常,用户不知道根据质量指标评估其测量的测量不确定性,尤其是在照度或亮度的现场光度测量中。在实现照明项目或现场其他特定测量之后,在照明系统的验证测量中会出现这种歧义。本文关注由标准草案CIE DS 023 / E:2012定义的不同类别照度计对特定照度现场测量的测量不确定度分析。研究工作的主要目的是在众所周知的条件下对室内工作场所进行现场测量。已知根据EN 12464-1室内工作场所的照明系统参数(即平均维持照度E_m,均匀度U_0等)和所测量房间中所用光源的光谱(在灯具中),可以将测量结果与所述不确定性进行比较假设正态分布,则对于95%的置信区间(k = 2)。假设标准草案中规定了质量指标,针对从较差等级到最佳质量等级的不同类型的照度计,研究了在该特定现场测量中的测量不确定度。从这项工作的结果可以在一定程度上有助于预测具体级别的二手照度计的测量准确性,从而避免使用它们进行现场测量的用户产生歧义。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号