This paper presents results of in vitro and in vivo trials of a wearable artificial kidney (WAK) based on peritoneal dialysis. Materials and methods: in vitro trials were carried out in a developed test bench, which imitates patient's peritoneal cavity and capable to measure main procedure parameters. In vivo trial was carried out on a 15 kg dog in a healthy state to evaluate the effect of WAK on blood biochemical indicators and in an acute kidney injury state to test WAK's blood purification capabilities. Results: in vitro trials demonstrated that WAK could eliminate urea, uric acid and creatinine from spent dialysate with mean mass rates of 0.85±0.1 g/h, 0.10±0.04 g/h and 0.05±0.01 g/h respectively. Concentrations of Na+, Cln-nand Can2+nions are kept in 10 % range from initial values. It is demonstrated that WAK can operate without replacement of expendable materials for 38 hours. During in vivo trial (39,5 hours), WAK performed continuous peritoneal dialysis with dialysis fluid regeneration. In acute kidney injury conditions (34 hours) creatinine and uric acid were eliminated from dialysing solution with the rate of 0,3 mg/h, urea was eliminated at the rate of 0,15 g/h, total removed ultrafiltrate volume was 350 ml. At the end of the second stage blood biochemical indicators stabilised in the range of normal values. Conclusion: developed WAK can perform elimination of metabolites from spent dialysis fluid with mass rates that are sufficient to maintain stable, physiologically normal metabolite concentrations in patient's blood.
机译:本文介绍了基于腹膜透析的可穿戴人工肾脏(WAK)的体外和体内试验结果。材料和方法:体外试验是在发达的试验台上进行的,该试验台模仿患者的腹膜腔并能够测量主要程序参数。在健康状态下对一只15公斤的狗进行了体内试验,以评估WAK对血液生化指标的影响,并在急性肾损伤状态下测试WAK的血液净化能力。结果:体外试验表明,WAK可以从废透析液中去除尿素,尿酸和肌酐,平均质量比分别为0.85±0.1 g / h,0.10±0.04 g / h和0.05±0.01 g / h。 Na +,Cln的浓度-< / sup> nand Can 2+ nions保持在初始值的10%范围内。事实证明,WAK可以运行38小时而无需更换消耗性材料。在体内试验(39.5小时)中,WAK进行了连续腹膜透析,并进行了透析液再生。在急性肾损伤情况下(34小时),透析液中肌酐和尿酸的清除率为0.3 mg / h,尿素的清除率为0.15 g / h,超滤液的总去除量为350 ml 。在第二阶段结束时,血液生化指标稳定在正常值范围内。结论:开发的WAK可以消除透析液用完后的代谢物,其质量速率足以维持患者血液中稳定,生理上正常的代谢物浓度。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号