The importance of modeling normal mammographic structure in optimizing flat-panel CT breast imaging systems

机译:在优化平板CT乳腺成像系统中建模正常的乳腺摄影结构的重要性

获取原文
获取原文并翻译 | 示例

摘要

In recent years, there has been interest in exploring the feasibility of CT breast imaging using flat-panel digital detectors in a truncated cone-beam geometry. Preliminary results are promising and it appears as if 3D tomographic imaging of the breast has great potential for reducing the masking effect of superimposed parenchymal structure typically observed with conventional mammography. In this study, a mathematical framework used for determining optimal design and acquisition parameters for such a CT breast imaging system is described. The ideal observer SNR is used as a figure-of-merit, under the assumptions that the imaging system is linear and shift-invariant. Computation of the ideal observer SNR used a parallel-cascade model to predict signal and noise propagation through the detector, as well as a realistic model of the lesion detection task in breast imaging. For all optimizations discussed here, the total mean glandular dose for a CT breast imaging study is constrained to be approximately equivalent to that of a two-view conventional mammography study. The framework presented is used to explore the affect of the specific task on the optimal exposure technique of flat-panel CT breast imaging. In particular, it is observed that modeling the normal mammographic structure in the projection images can sometimes impact the optimal kVp settings.
机译:近年来,人们一直在探索使用锥面几何形状的平板数字检测器进行CT乳腺成像的可行性。初步结果令人鼓舞,并且似乎乳房的3D断层扫描成像具有降低通常用常规X线照相术观察到的叠加实质结构的掩盖效果的巨大潜力。在这项研究中,描述了用于确定此类CT乳腺成像系统的最佳设计和采集参数的数学框架。在成像系统是线性且位移不变的假设下,理想观察者SNR用作品质因数。理想观察者SNR的计算使用了并行级联模型来预测信号和噪声通过检测器的传播,以及使用乳腺成像中病变检测任务的真实模型。对于此处讨论的所有优化,CT乳腺成像研究的总腺平均剂量被约束为近似等于两视图常规乳房X线照片研究的总腺剂量。提出的框架用于探讨特定任务对平板CT乳腺成像最佳曝光技术的影响。特别地,观察到在投影图像中对正常的乳房X线照相结构进行建模有时会影响最佳的kVp设置。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号