We have developed an instrument for real time continuous interferometric measurement of dimensional change in stimuli-responsive hydrogel sensors. The instrument consists of a superfluorecent LED and a low-cost grating spectrometer. All signal conditioning and processing is performed using a microprocessor. We have tested the instrument with a glucose sensitive hydrogel sensor. The rms noise on the length measurement is less than 1.4 nm, whereas rms noise on the length change is less than 0.25 nm. An observed bias of the length measurement can be reduced by minor changes in the frequency estimation algorithm used.
机译:我们已经开发了一种用于对刺激响应水凝胶传感器进行尺寸变化的实时连续干涉测量的仪器。该仪器包括一个超荧光LED和一个低成本的光栅光谱仪。所有信号调节和处理均使用微处理器执行。我们已经使用葡萄糖敏感性水凝胶传感器对仪器进行了测试。长度测量的均方根噪声小于1.4 nm,而长度变化的均方根噪声小于0.25 nm。可以通过使用的频率估计算法中的微小变化来减少观察到的长度测量偏差。