Simulation of pseudo-clinical conditions and image quality evaluation of PET scanner based on pixelated CdTe detector

机译:基于像素化CdTe检测器的PET扫描仪伪临床条件模拟和图像质量评估

获取原文

摘要

A novel conceptual design of PET scanner, known as a Voxel Imaging PET (VIP) [1], based on pixelated solid state detector, has been proposed to overcome the intrinsic limitations of state-of-art PET devices based on scintillating crystals. For this study, the VIP scanner has been simulated in different pseudo-clinical conditions in order to assess its image quality performance. The evaluation of the PET image quality follows the NEMA prescriptions. Results are also presented for both analytic and iterative image reconstruction methods. The preliminary results from simulations show that the image reconstruction performance of the VIP scanner under realistic clinical conditions provides good images with low dose. The VIP scanner is able to detect down to 1 mm in diameter hot regions in small phantoms without background activity and down to 5 mm in diameter hot spheres in big phantoms in the presence of background activity. Moreover, good quality and high contrast images can be obtained with a considerably lower number of coincidences with respect to the usual crystal PETs.
机译:已经提出了一种基于像素化固态检测器的新颖的PET扫描仪概念设计,称为Voxel Imaging PET(VIP)[1],以克服基于闪烁晶体的最新PET设备的固有局限性。对于本研究,已在不同的伪临床条件下模拟了VIP扫描仪,以评​​估其图像质量性能。 PET图像质量的评估遵循NEMA规定。还提供了分析图像和迭代图像重建方法的结果。模拟的初步结果表明,VIP扫描仪在实际临床条件下的图像重建性能可提供低剂量的良好图像。 VIP扫描仪能够在没有背景活动的情况下检测小体模中直径低至1毫米的高温区域,而在有背景活动的情况下检测大体模中直径低至5毫米的热球。而且,相对于通常的晶体PET,可以以相当低的重合次数获得高质量和高对比度的图像。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号