【24h】

Normalization of Monte Carlo PET data using GATE

机译:使用GATE对Monte Carlo PET数据进行归一化

获取原文

摘要

The reconstruction of PET data without proper correction for non-uniform detection efficiency leads to severe artifacts in the image. This is also true for simulated data, particularly for Monte Carlo simulations that aim to realistically model the detectors. In the present study, we showed that the simulation with GATE software of the ECAT HR+, a block detector ring scanner, results in non negligible variations across the lines of response in the detection sensitivity. Like for real scanners, the line of response efficiency depends on its angle with the crystal axis and the intrinsic crystal efficiencies inside a block.
机译:如果没有对不均匀的检测效率进行适当的校正,则重建PET数据会导致图像中出现严重的伪影。对于模拟数据也是如此,尤其是对于旨在对检测器进行实际建模的蒙特卡洛模拟。在本研究中,我们显示了使用ECAT HR +(块检测器环形扫描仪)的GATE软件进行的仿真导致整个检测灵敏度范围内响应线的变化均不可忽略。与实际扫描仪一样,响应效率线取决于其与晶轴的夹角以及模块内部固有的晶体效率。