There have been increasing reports of wear and corrosion of modular junctions between the femoral balls and stems. The assessment of modular component corrosion has been largely based on the subjective, visual inspection of explanted modular interfaces. In this study, 51 modular specimens from a variety of total hip explants were given visual corrosion scores by three observers, first after a light rinsing (R1) and then again after a thorough soaking and scrubbing (R2), to determine the influence of cleaning on visual scoring methods. Each observer scored the specimens three separate times, and each time, two different scoring methods were used: the four-point Goldberg method (G) and the five-point Fricka method (F). After thoroughly cleaning the specimens, the intraobserver reliability of both methods increased (6: R1-0.62, R2-0.70; F: R1-0.50, R2-0.64). Although the interobserver reliability for the Goldberg method was unchanged (R1-0.54, R2-0.55), the interobserver reliability of the Fricka method substantially decreased (R1-0.41, R2-0.15), suggesting that cleaning lowered the consistency in methodology for this scoring method. Additionally, the reliability between the first and second round was found to be higher for the Goldberg method than for the Fricka method (G-0.62, F-0.47). Differences between first-round and second-round scores ranged from 0 to 2 points. Both methods emphasize the dark discoloration, although some may be artifactual; therefore, cleaning is recommended to remove any debris and to expose surface features. In the absence of deposits and evidence of fretting, it was hard to assign an appropriate score in cases in which the predominant feature was surface grooving, which was not included in the criteria for either method. Visual corrosion scoring would be improved if the score accounted for the formation of grooves that are a common feature of many explanted tapers.
机译:越来越多的关于股球和股骨之间的模块连接处的磨损和腐蚀的报道。模块化组件腐蚀的评估主要基于对植入式模块化接口的主观,目视检查。在这项研究中,三位观察者对来自各种全髋关节外植体的51个标本进行了视觉腐蚀评分,首先是在轻冲洗(R1)之后,然后是在彻底浸泡和擦洗(R2)之后,以确定清洁的影响视觉评分方法。每个观察者分别对样本进行3次评分,并且每次都使用两种不同的评分方法:四点Goldberg方法(G)和五点Fricka方法(F)。彻底清洁样品后,两种方法的观察者内部可靠性均得到提高(6:R1-0.62,R2-0.70; F:R1-0.50,R2-0.64)。尽管Goldberg方法的观察者间可靠性不变(R1-0.54,R2-0.55),但Fricka方法的观察者间可靠性大大降低(R1-0.41,R2-0.15),这表明清洗降低了该评分方法的一致性方法。此外,发现Goldberg方法的第一轮和第二轮之间的可靠性高于Fricka方法的可靠性(G-0.62,F-0.47)。第一轮和第二轮得分之间的差异为0到2分。两种方法都强调深色,尽管有些可能是人为的。因此,建议清洁以清除所有碎屑并暴露表面特征。在没有沉积物和微动的迹象的情况下,如果主要特征是表面开槽,则很难分配适当的分数,这两种方法的标准均未包括在内。如果该分数说明了许多外植锥度的共同特征,那就是凹槽的形成,则可以改善视觉腐蚀评分。

著录项

  • 来源
  • 会议地点 New Orleans LA(US)
  • 作者单位

    UCLA, Orthopaedic Biomechanics and Mechanobiology Research Lab, Orthopaedic Institute for Children, 403 West Adams Blvd., Los Angeles, CA 90007,UCLA, Dept. of Bioengineering, Los Angeles, CA 90007;

    UCLA, Orthopaedic Biomechanics and Mechanobiology Research Lab, Orthopaedic Institute for Children, 403 West Adams Blvd., Los Angeles, CA 90007,UCLA, Dept. of Orthopaedic Surgery, Los Angeles, CA 90007;

    UCLA, Orthopaedic Biomechanics and Mechanobiology Research Lab, Orthopaedic Institute for Children, 403 West Adams Blvd., Los Angeles, CA 90007,UCLA, Dept. of Orthopaedic Surgery, Los Angeles, CA 90007;

    UCLA, Orthopaedic Biomechanics and Mechanobiology Research Lab, Orthopaedic Institute for Children, 403 West Adams Blvd., Los Angeles, CA 90007,UCLA, Dept. of Orthopaedic Surgery, Los Angeles, CA 90007;

  • 会议组织
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

    arthroplasty; corrosion; retrieval; taper;

    机译:关节置换术腐蚀;恢复;锥度;
  • 入库时间 2022-08-26 14:20:13

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号