首页> 外文会议> >A comparative analysis of the #x2018;transmittance#x2019; method between using integrating sphere and filter holders for measuring light absorption by aquatic particles
【24h】

A comparative analysis of the #x2018;transmittance#x2019; method between using integrating sphere and filter holders for measuring light absorption by aquatic particles

机译:使用积分球和滤光片支架测量水生颗粒的光吸收率的“透射率”方法的比较分析

获取原文

摘要

Measuring light absorption of suspended matter in water is essential for water color remote sensing. The ‘transmittance’ method, which is based on the Quantitative Filter Technology, is one of the most widely used methods to determine absorption of suspended matters in water. This method measures optical density of suspended matters that concentrated on the filter and then calculate absorption by applying light pathlength amplification correction to the optical density. Either a pair of filter holders or an integrating sphere is the necessary accessory for the measurement of the optical density. However, according to our research, we discovered that the results obtained with different accessories have some differences. By comparing the optical density values that measured with different accessories, we analyzed the difference between different accessories and proposed a revised model for light pathlength amplification correction.
机译:测量水中悬浮物的吸光度对于彩色遥感至关重要。基于定量过滤技术的“透射率”方法是确定水中悬浮物吸收率的最广泛使用的方法之一。该方法测量聚集在滤光片上的悬浮物的光密度,然后通过对光密度应用光程长度放大校正来计算吸收。一对滤光片架或一个积分球是测量光密度的必要附件。但是,根据我们的研究,我们发现使用不同附件获得的结果存在一些差异。通过比较使用不同附件测量的光密度值,我们分析了不同附件之间的差异,并提出了用于光程长度放大校正的修订模型。