Impact of Digitalization of Mammographic Units on Average Glandular Doses in the Flemish Breast Cancer Screening Program

机译:乳腺癌单位数字化对佛兰氏乳腺癌筛查计划平均腺体剂量的影响

获取原文

摘要

The impact of digitalization on the average glandular doses in 49 mammographic units participating in the Flemish Breast Cancer Screening Program was studied. Screen-film was changed to direct digital radiography and computed radiography in 25 and 24 departments respectively. Average glandular doses were calculated before and after digitalization for different PMMA-phantom thicknesses and for groups of 50 successive patients. For the transition from screen-film to computed radiography both phantom and patient dose data show a significant increase of dose with digitalization. For the transition from screen-film to direct digital radiography the evolution of the average glandular dose depends on the phantom thickness. For 20mm PMMA a significant increase in dose was found, for 45mm and 70mm PMMA there was a significant decrease in dose. The median average glandular dose of the patient dosimetry showed a smaller but significant decrease.
机译:研究了参与佛兰氏乳腺癌筛查计划的49个乳腺癌平均腺剂量对平均腺剂量的影响。屏幕薄膜分别改变为25和24个部门的直接数字射线照相和计算的射线照相。在数字化之前和之后计算平均腺剂量,用于不同的PMMA - 幻影厚度和50名连续患者的组。对于从屏幕膜到计算的射线照相的过渡,幻影和患者剂量数据都显示出具有数字化的剂量的显着增加。对于从筛膜到直接数字射线照相的转变,平均腺体剂量的演变取决于幻影厚度。对于20mm PMMA,发现剂量显着增加,45mm和70mmMMA的剂量显着降低。患者剂量率的中值平均腺剂量显示出较小但显着降低。