Effects of Window Width and Window Level Adjustment on Detection tasks in Computed Tomography Images

机译:窗口宽度和窗口电平调整对计算机断层扫描图像中检测任务的影响

获取原文

摘要

This study experimentally evaluated the effect of window width (WW) and window level (WL) on the task based detectability index (d'). Window-level transformation is frequently performed on CT images to improve visualization in clinical arena. Numerous model observers and metrics have been used to assess CT image quality. However, objective assessment is typically performed on the reconstructed CT image without considering the WW and WL settings used by the reader. In this study, the ACR CT phantom was scanned at 120 kV and 90 mAs and images were reconstructed using filtered backprojection. The bone and acrylic contrast objects from module one of the ACR phantom were selected for calculating the effect of the WW/WL on the detectability index (d'). The d' for each object at 90 mAs was calculated for a range of WW and WL values. The results demonstrated that the d' values were affected by the WW and WL settings. For example, WL setting of 20 HU and window width of 150 HU resulted in a 35% decrease in d' compared to that of the untransformed image. A WL of 20 and WW of 1400 resulted in a 33% decrease in the d' value for the high contrast object. The investigated WW and WL settings did not improve d' for any of the investigated objects when reconstructed with the standard kernel. The results suggest that d' is affected by the WW/WL settings. However, because d' does not model the contrast adaptation of the human visual system, it may not represent the change in perceived image quality with window-level transformation.
机译:本研究通过实验评估了窗宽(WW)和窗口电平(WL)对基于任务的可检测性索引(D')的影响。在CT图像上经常执行窗口级变换,以改善临床竞技场中的可视化。许多模型观察者和指标已经用于评估CT图像质量。然而,通常对重建的CT图像进行客观评估,而不考虑读取器使用的WW和WL设置。在该研究中,将ACR CT幻影在120kV下扫描,并使用过滤的反射重建90μm和图像。选择来自模块的骨骼和丙烯酸腺对比度对比物,用于计算WW / WL对可检测性指数(D')的影响。计算90 mas的每个对象的d'是计算的一系列WW和WL值。结果表明,D'值受WW和WL设置的影响。例如,与未转化的图像相比,20 HU和窗宽的WL设置为150U的窗口宽度导致D'下降35%。对于高对比度对象的D'值,WL为20和WW的WL导致33%的减少。在用标准内核重建时,调查的WW和WL设置没有改善任何调查的对象。结果表明D'受WW / WL设置的影响。然而,因为D'不模型人类视觉系统的对比度适应,它可能不代表具有窗口级变换的感知图像质量的变化。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号