An approach is presented to cancel the arterial wall contamination using audio Doppler ultrasound signals from two consecutive cardiac cycles. The wall signal is approximately periodic, while the Doppler signal from the blood flow is random because the scatters within the ultrasound sample volume are continually updated. Thus, the signal subtraction between two consecutive cardiac cycles is carried out in time domain. Simulated Doppler signals from a normal femoral artery with various clutter to blood ratios are used to validate this approach. The conventional high-pass filtering method is also carried out for comparison. The experimental results show that this approach may be a good solution for the wall thump removal when there is few difference of the wall movement between the two consecutive cardiac cycles.
机译:提出一种使用来自两个连续心脏周期的音频多普勒超声信号取消动脉壁污染。 壁信号大致周期性,而来自血流的多普勒信号是随机的,因为超声样本体积内的散射不断更新。 因此,在时域中执行两个连续的心脏周期之间的信号减法。 使用各种杂波与血液比率的常规股动脉的模拟多普勒信号用于验证这种方法。 还进行了传统的高通滤波方法以进行比较。 实验结果表明,当两个连续的心脏周期之间的壁运动差异很小时,这种方法可以是壁堆块去除的良好解决方案。