首页> 外文会议>Annual Meeting of the Institute of Nuclear Materials Management >Calibration of Active Well Coincidence Counter (AWCC) with Highly Enriched Uranium (HEU) and AmBe Active Neutron Interrogation Source at the Savannah River National Laboratory
【24h】

Calibration of Active Well Coincidence Counter (AWCC) with Highly Enriched Uranium (HEU) and AmBe Active Neutron Interrogation Source at the Savannah River National Laboratory

机译:在萨凡纳河国家实验室用高浓缩铀(HEU)和AmBe主动中子探测仪源校准主动井符合计数器(AWCC)

获取原文

摘要

The Active Well Coincidence Counter (AWCC) is a neutron multiplicity counter developed at Los Alamos National Laboratory (LANL) and used primarily to verify total fissile mass (~(235)U + ~(239)Pu) present in a sample of special nuclear material (SNM). The AWCC is comprised of a sample well surrounded by the two rings of ~3He detector tubes aligned in a vertical axis to measure neutrons emitted from SNM samples. Each of the ring consists of 21 detector tubes housed inside high-density polyethylene (HDPE) moderator wells. The AWCC was designed to use two AmLi fixed neutron sources in active interrogation mode. Since these sources can be difficult to acquire, experimental and MCNP studies were performed at the Nuclear Measurements group at the Savannah River National Laboratory (SRNL) to calibrate the AWCC with AmLi and AmBe active neutron sources. The objective of this work was to obtain an alternative active neutron source for this nondestructive assay (NDA) instrument and possibly for other NDA instruments that rely on traditional AmLi neutron sources. AmLi and AmBe active background rates, induced fission (IF) singles and doubles rates, and IF doubles to singles ratio were compared. The instrument was calibrated in MCNP for 93% enriched HEU samples of different density. Finally, effects of sample geometry to the IF count-rates was also investigated. Results showed that the AmBe source had a higher active singles background than the AmLi source (15% versus 5%). AmBe also had a higher active doubles background count rate (8.647 cps versus 0 cps). In addition, AmBe had a lower net IF singles and doubles rate than AmLi (by approximately 52% and 31% respectively). Even though AmBe provided higher active singles and doubles background rates and lower net IF singles and doubles rates, AmLi resulted in lower net doubles to net singles ratio (-40%) compared to AmBe. This shows AmBe can be used in lieu of AmLi in the AWCC instrument.
机译:主动井符合计数器(AWCC)是洛斯阿拉莫斯国家实验室(LANL)开发的中子多重性计数器,主要用于验证特殊核材料(SNM)样品中的总裂变质量(~(235)U+~(239)Pu)。AWCC由一个样品井组成,周围环绕着两圈~3He探测管,这些探测管沿纵轴排列,用于测量SNM样品发射的中子。每个环由21根探测管组成,位于高密度聚乙烯(HDPE)慢化剂井内。AWCC设计用于在主动询问模式下使用两个AmLi固定中子源。由于这些源可能难以获得,萨凡纳河国家实验室(SRNL)的核测量组进行了实验和MCNP研究,以使用AmLi和AmBe活性中子源校准AWCC。这项工作的目的是为该无损检测(NDA)仪器以及依赖传统AmLi中子源的其他NDA仪器获得替代活性中子源。比较AmLi和AmBe活动背景率、诱发裂变(IF)单体和双体率以及IF双体与单体比率。该仪器在MCNP中校准,用于不同密度的93%浓缩高浓缩铀样品。最后,还研究了样品几何形状对IF计数率的影响。结果表明,AmBe源比AmLi源具有更高的活跃单身背景(15%对5%)。AmBe的活跃双打背景计数率也较高(8.647次/秒,而非0次/秒)。此外,AmBe的净IF单打和双打率低于AmLi(分别约为52%和31%)。尽管AmBe提供了更高的活跃单打和双打背景率,以及更低的净IF单打和双打率,但与AmBe相比,AmLi导致了更低的净双打与净单打比率(-40%)。这表明AmBe可以代替AWCC仪器中的AmLi。

著录项

相似文献

  • 外文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号