Design and Calibration of a Polychromator for the Thomson Scattering at Wendelstein 7-X

机译:WENDELSTEIN 7-X的汤姆森散射的多彩器的设计与校准

获取原文

摘要

On Wendelstein 7-X, electron density and electron temperature profiles are planned to be measured by Thomson scattering. The specific system employs fiber bundles for the scattered light which is detected by means of a five-channel polychromator system. The detection system is assembled and its components are characterized. A calibration technique employing a super-continuum light source is assessed by comparison with a calibration technique using an optical parametric oscillator laser system. The characterization is preparing further diagnostic optimization studies employing a virtual instrument.
机译:在Wendelstein 7-x上,计划通过汤逊散射测量电子密度和电子温度曲线。特定系统采用用于散射光的纤维束,该散射光通过五通道多彩器系统检测。组装检测系统,其部件的特征在于。通过使用光学参数振荡器激光系统的校准技术进行评估采用超连体光源的校准技术。表征正在制备采用虚拟仪器的进一步诊断优化研究。