首页> 外文会议>Conference on Space Telescopes and Instrumentation: Ultraviolet to Gamma Ray >Development and performance verification of a TPC polarimeter for high energy X-rays
【24h】

Development and performance verification of a TPC polarimeter for high energy X-rays

机译:高能量X射线TPC偏振仪的开发与性能验证

获取原文

摘要

Wo present the first application of a time projeetion chamber polarimeter to measure high energy X-ray polarization above 10 keV. The polarimeter is designed based on the PRAXyS soft X-ray polarimeter. The scaled gas is changed to a gas mixture of 60% argon and 40% dimethyl ether at 1 atm to be sensitive to high energy X-rays. The polarimeter performance is verified with linearly polarized, monochromatic X-rays at a synchrotron radiation facility, KEK Photon Factory BL-14A. The measured modulation factors are 42.4 ±0.6%, 50.4 ±0.6%, and 55.0 ± 0.6% at 12, 14, and 16 keV, respectively, and the measured polarization angles arc consistent with the expected values at all energies.
机译:WO呈现了时间工程室偏振仪的第一次应用,以测量高于10keV的高能X射线极化。偏振仪基于Praxys软X射线偏振仪设计。将缩放的气体改变为60%氩气的气体混合物和40%二甲醚,在1ata下对高能X射线敏感。在Synchrotron辐射设施,KEK Photon Factory BL-14A中,通过线性偏振,单色X射线验证了极性仪性能。测量的调制因子分别为12,14和16kev的42.4±0.6%,50.4±0.6%和55.0±0.6%,测量的偏振角恒定与所有能量的预期值一致。