Dental and maxillofacial cone beam computed tomography absorbed dose distribution calculation by GEANT4

机译:牙科和颌面锥梁计算机断层扫描通过GEANT4吸收了剂量分布计算

获取原文

摘要

Absorbed dose distributions in dental and maxillofacial cone beam computed tomography (dental CBCT) are essential to dental CBCT dose indices. Direct measurements by thermoluminescence detectors are laborious. We establish a valid GEANT4 based absorbed dose simulation program with mean deviation being 7.25% compared to experimental measurements. Dental CBCT absorbed dose distributions simulated by this program indicate: The distributions are not always symmetry; Non-symmetry cases are when phantom center is departured from isocenter, half-fan beam with 360° scan angle range and some case full-fan beam with less than 360° scan angle range; Dose index weights for circular symmetry absorbed dose distributions are quite different for fields of view diameter bigger and smaller than phantom diameter.
机译:牙科和颌面锥梁的吸收剂量分布计算机断层扫描(牙科CBCT)对牙科CBCT剂量指数至关重要。热致发光探测器的直接测量是费力的。与实验测量相比,我们建立了基于Geant4的吸收剂量仿真程序,其平均偏差为7.25%。通过该程序模拟的牙科CBCT吸收剂量分布表明:分布并不总是对称;非对称性情况是当幻象中心脱离Isocenter,半风扇梁,带有360°扫描角度范围的半风扇梁和一些壳体扇形扫描角范围小于360°扫描角度范围;对于圆对称吸收剂量分布的剂量指数重量对于较大且小于幻像直径的视野而异。