Measurement of X-ray Beam Quality and Patient Dose Information from Analog Mammography

机译:从模拟乳房X线摄影测量X射线光束质量和患者剂量信息

获取原文

摘要

Measurement of X-ray beam quality and patient dose information from analog mammography have been studied by using an X-ray multimeter from 2016-2018.Parameters in the study are voltage accuracy, reproducibility test, radiation output linearity, radiation beam quality test, and patient dose information.The patient dose information using PMMA phantom with a thickness 4 cm in the Mean Glandular Dose(MGD)obtained dose values in 2016, 2017, 2018 are 0.982 mGy, 0.982 mGy, 0.953mGy, respectively.This study aims to determine the feasibility of mammography, while patient doses are compared with the standars set by the IAEA.The results showed that the suitability test of X-ray beam quality and patient dose information still comply the requirements recommended by the IAEA.
机译:通过使用2016-2018的X射线万用表研究了模拟乳房X线摄影X射线光束质量和患者剂量信息的测量。研究中的参数是电压精度,再现性测试,辐射输出线性,辐射光束质量测试,以及患者剂量信息。患者剂量信息在平均腺剂量(MGD)中使用厚度为4厘米的PMMA幻影(MGD)获得了2016年的剂量值,分别为0.982 MGY,0.982 MGY,0.953吨。该研究旨在确定乳房X线照相术的可行性,而患者剂量与由原子能机构设定的静物相比。结果表明,X射线束质量和患者剂量信息的适用性试验仍然符合IAEA推荐的要求。