Several technical approaches were evaluated to improve the performance of rescuers, but to our knowledge no attempt was realized to develop a feedback device to supervise the combination of passive ventilation and chest compressions. The purpose of this work is the design and the implementation of a new CPR feedback device, based on an accelerometer to measure chest compressions (CC) depth and frequency, flow and pressure sensors ensuring the measurement of the passive ventilation generated by only CC and an user interface installed on a computer to acquire and supervise, in real time, the different variations. The proposed device is tested by 20 volunteers inexperienced in CPR with 2 minutes of continuous CC on a thoracic lung model, which we developed, to evaluate the functioning of the device and observe its effect on the participant's performance. The 2 min mean CC rate was within AHA 2015 guidelines with the use of the device, where 74% of CC frequency was in adequate range. Correct CC depth was achieved by 53 % of all compressions with a mean of 52.6 ± 7.7 mm. Correct CC depth was greater for male vs. female (p = 0.002). However, CC frequency rate was greater for female vs male (p=0.003). 45 % of CC are performed in correct rate (100-120 min-1) and correct depth (50-60 mm) using the device. The intra-thoracic pressure generated by the CC alone was 18.9 ± 3.9 cmH2O with a tidal volume of 77.2 ± 19.6 ml. Our results are in accordance with previous findings. Further clinical studies are required in order to evaluate its efficiency and see if it contributes to an improvement in the CPR outcomes.
机译:评估了几种技术方法以提高救援人员的性能,但对于我们的知识,没有意识到开发反馈装置以监督被动通风和胸部按压的组合。这项工作的目的是设计和实施新的CPR反馈装置,基于加速度计测量胸部按压(CC)深度和频率,流量和压力传感器,确保仅由CC和AN产生的被动通风的测量用户界面安装在计算机上以实时获取和监督不同的变体。所提出的设备由20名志愿者进行测试,在CPR中缺乏经验,在我们开发的胸肺模型上连续CC,以评估设备的运作并观察其对参与者的绩效的影响。 2分钟的平均CC速率在AHA 2015的指导方针中使用了该装置,其中74%的CC频率是足够的范围。通过53 %的所有压缩来实现正确的CC深度,平均为52.6±7.7mm。对于雌性对雌性的正确CC深度较大(P = 0.002)。然而,对于雌性VS雄性的CC频率率更大(P = 0.003)。 45 %的CC以正确的速率进行(100-120分钟 -1 )使用该装置的正确深度(50-60毫米)。 CC单独产生的胸内压力为18.9±3.9 CMH 2 o潮量为77.2±19.6 ml。我们的结果符合以前的研究结果。需要进一步的临床研究,以评估其效率,并看看它是否有助于改善CPR结果。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号