Computer Tomography Tube Voltage and Phantom Dimensions Influence on the Number of Hounsfield Units

机译:计算机断层扫描管电压和幻象尺寸对Hounsfield单位数量的影响

获取原文

摘要

Hounsfield units (HU) are a dimensionless units universally used in computed tomography (CT) scanning to express CT numbers in a standardized and convenient form. Calibration curve (CC) implemented in radiotherapy treatment planning system (TPS) shows the dependence of HU on relative electron density (RED). Radiotherapy treatment planning process is based on data from the CC and linear accelerator (Linac) data set. RED is constant and specified for each material, but the HU can change depending on several factors. Dosimetric phantom in radiotherapy properly represents human anatomy and enable thorough analysis of both the imaging and dosimetry system. For the purpose of CC checking, two phantoms were used: Catphan 504 (The Phantom Laboratory) and CIRS 062MA (Computerized Imaging reference Systems, Inc.). The aim of the paper was to check whether the HU depends on the dimensions of the phantom. Catphan 504 phantom was scanned first, in order to verify the HU determination methodology. After that, CIRS 062MA phantom representing both abdomen and head configuration was scanned. By analyzing Catphan 504 phantom obtained scanning results, it was concluded that the HU values are correctly determined, since the results are in accordance with the manufacturer's recommendations. Based on results of CIRS 062MA phantom scan, the HU dependence on the CT simulator tube voltage was confirmed. Significant difference in HU for high density materials (up to 135 HU) due to different phantom sizes was observed.
机译:Hounsfield Units(Hu)是一种普遍用于计算机断层扫描(CT)扫描的无量纲单元,以以标准化和方便的形式表达CT号。放射疗法治疗系统(TPS)中实施的校准曲线(CC)显示了HU对相对电子密度(红色)的依赖性。放射疗法治疗计划过程基于来自CC和线性加速器(LINAC)数据集的数据。红色是恒定的,为每种材料指定,但HU可以根据若干因素而改变。放射疗法中的剂量幽灵正常代表人类解剖学,并能够彻底分析成像和剂量测量系统。出于CC检查的目的,使用了两种幽灵:CATPHAN 504(PHANTOM实验室)和CIRS 062mA(计算机成像参考系统,INC。)。本文的目的是检查胡是否取决于幻象的尺寸。第一次扫描Catphan 504幻影,以验证HU确定方法。之后,扫描表示腹部和头部配置的CIRS 062mA幻影。通过分析CATPHAN 504幻影获得的扫描结果,得出结论是,HU值是正确确定的,因为结果符合制造商的建议。基于CIRS 062mA幻影扫描的结果,确认了对CT模拟器管电压的HU依赖性。由于观察到由于不同的体模尺寸,HU对于高密度材料(高达135U的HU)的显着差异。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号