首页> 外文会议>International Nuclear Science, Technology and Engineering Conference >Dosimetry of Al2O3 optically stimulated luminescent dosimeter at high energy photons and electrons
【24h】

Dosimetry of Al2O3 optically stimulated luminescent dosimeter at high energy photons and electrons

机译:高能量光子和电子的Al2O3光学刺激发光剂量计的剂量法

获取原文

摘要

The linearity of Al2O3 OSL dosimeters (OSLD) were evaluated for dosimetry works in clinical photons and electrons. The measurements were made at a reference depth of Z_(ref) according to IAEA TRS 398:2000 codes of practice at 6 and 10 MV photons and 6 and 9 MeV electrons. The measured dose was compared to the thermoluminescence dosimeters (TLD) and ionization chamber commonly used for dosimetry works for higher energy photons and electrons. The results showed that the measured dose in OSL dosimeters were in good agreement with the reported by the ionization chamber in both high energy photons and electrons. A reproducibility test also reported excellent consistency of readings with the OSL at similar energy levels. The overall results confirmed the suitability of OSL dosimeters for dosimetry works involving high energy photons and electrons in radiotherapy.
机译:在临床光子和电子中评估了Al2O3 OSL剂量计(OSLD)的线性度。根据IAEA TRS 398:2000在6和10mV光子和6和9MEV电子的实施实践码的Z_(REF)的参考深度下进行测量。将测量剂量与热致敏剂剂量计(TLD)和电离室进行比较,常用于剂量测定作品,用于更高的能量光子和电子。结果表明,OSL剂量计中的测量剂量与高能光子和电子的电离室报道良好。再现性测试还报告了在类似能量水平的OSL中读数的优异一致性。总体结果证实了OSL剂量计的适用性涉及高能光子和放射疗法中的电子的剂量测量仪。