Certain measurable parameters may alter laboratory and pharmaceutical quality control tests. In order to prove such statement, the following parameters were monitored: samples and reagents storage temperature, distilled water quality and electrical noise. The results obtained showed that the centers under analysis do not comply with the infrastructure local rules for the installment of the related equipment and that they neither comply with the storage specifications provided by the suppliers of the reagents. Key Words- Storage of samples and reagents, distilled water, electrical noise.
机译:某些可测量的参数可以改变实验室和药物质量控制测试。为了证明这种陈述,监测以下参数:样品和试剂储存温度,蒸馏水质量和电噪声。所获得的结果表明,分析中的中心不符合基础设施本地规则,以分配相关设备,并且它们既不符合试剂供应商提供的存储规范。关键词 - 样品和试剂的储存,蒸馏水,电噪声。