首页> 外文会议>International Conference on Sensors, Mechatronics and Automation >A Simulation Study for the attenuation correction of Respiratory Motion with the PET/CT Gating data
【24h】

A Simulation Study for the attenuation correction of Respiratory Motion with the PET/CT Gating data

机译:宠物/ CT门控数据呼吸运动衰减校正的仿真研究

获取原文

摘要

Objective To investigate the effect of attenuation correction on the gating data based on the true coincidences event of detector ring with the application of PET/CT. Methods Use the software GATE that can simulate the imaging of PET/CT to do the pixel phantom simulation in order to simulate the respiratory motion of the lung in the process of PET/CT scanning. Then, do the gating process and the attenuation correction with the simulation data of GATE. Results The quality of gating images with attenuation correction is better than none attenuated images. Conclusion The quality of gating images with attenuation correction can further improve the quality of images and the accuracy of diagnosis.
机译:目的探讨探测器圈与PET / CT的真正巧合事件的衰减校正对门控数据的影响。 方法使用能够模拟PET / CT的成像的软件门,以做像素幻像仿真,以便在PET / CT扫描过程中模拟肺的呼吸运动。 然后,使用门的模拟数据进行门控过程和衰减校正。 结果具有衰减校正的门控图像的质量优于没有衰减图像。 结论具有衰减校正的门控图像的质量可以进一步提高图像的质量和诊断的准确性。