首页> 外文会议>International Conference on Mechanical, Electronic and Engineering Technology >Experimental Studies of the Longitudinal Flexibility of Balloon-expandable Coronary Stents
【24h】

Experimental Studies of the Longitudinal Flexibility of Balloon-expandable Coronary Stents

机译:气囊可膨胀冠状动用支架纵向灵活性的实验研究

获取原文

摘要

Stents are medical devices used in cardiovascular intervention for unblocking the diseased arteries and restoring blood flow. A setup for the measurement of the longitudinal flexibility of a coronary stent was developed based on machine vision technology. A crimped stent made by medical stainless steel thin-walled tube was tested with the setup. The results show that the bending deformation of stent includes two phases: the elastic deformation phase and the plastic deformation phase. The bending angle changed very little with the increase of bending moment during the elastic deformation. However, the bending angle changed significantly when a small moment was applied during the plastic deformation. In conclusion, the experimental setup can be used to study and compare flexibility of different design kinds of coronary stents and provides a convenient tool for designers to improve bending characteristics of new stents.
机译:支架是用于脱离患病动脉和恢复血流的心血管干预的医疗装置。 基于机器视觉技术开发了测量冠状动脉支架纵向灵活性的设置。 用设定测试由医用不锈钢薄壁管制成的压接支架。 结果表明支架的弯曲变形包括两个阶段:弹性变形相和塑性变形阶段。 随着弹性变形期间弯矩的增加,弯曲角变化很小。 然而,当在塑性变形期间施加小时刻时,弯曲角显着变化。 总之,实验装置可用于研究和比较不同设计冠状动用支架的灵活性,为设计人员提供了一种方便的工具,以改善新支架的弯曲特性。