PRIMARY CALIBRATION OF SHOCK TRANSDUCERS LASER POSITION INFLUENCES MEASUREMENT RESULTS

机译:震动传感器的主要校准激光位置影响测量结果

获取原文

摘要

For the primary calibration of shock transducers according to ISO 16063-13 on a calibration system with Hopkinson-bar shock exciter a laser vibrometer is used as a reference sensor. Earlier publications have shown that the waves in the Hopkinson-Bar seem not to propagate exactly as plain waves and that it thus matters where the laser spot is positioned on the surface of the bar. However, in these investigations the shock duration could not be changed in a controlled way. This paper presents results from measurements by means of a special Hopkinson-bar that allows to vary the shock duration precisely and thus allows to measure the frequency dependency of such effects.
机译:对于根据ISO 16063-13在带有Hopkinson-Bar冲击激励器的校准系统上的冲击换能器的主要校准,激光振动计用作参考传感器。早期的出版物表明,Hopkinson-Bar中的波似乎不完全像普通波一样传播,因此激光光斑定位在杆的表面上的情况。但是,在这些调查中,震荡持续时间无法以受控方式改变。本文通过特殊的HopKinson-Bar提供了测量结果,允许精确地改变震动持续时间,从而允许测量这些效果的频率依赖性。