Calibrating Image Plate Sensitivity in the 700 to 5000 eV Spectral Energy Range

机译:校准图像板敏感度在700至5000eV光谱能量范围内

获取原文

摘要

This paper describes a method to calibrate image plate sensitivity for use in the low energy spectral range. Image plates, also known as photostimulable luminescence (PSL) detectors, have often proved to be a valuable tool as a detector for plasma physics studies. Their advantages of large dynamic range, high stopping power, and resistance to neutron damage sometimes outweigh the problems of limited resolution and the remote processing required. The neutron damage resistance is required when the X-ray source is producing a high neutron flux. The Static X-ray Imager (SXI) is a key diagnostic on the National Ignition Facility (NIF) target chamber at LLNL for use in determining the symmetry of the laser beams. The SXI is essential to proper interpretation of the data from the Dante diagnostic to determine the X-ray radiation temperature1. It is comprised of two diagnostics located at the top and the bottom of the target chamber. The usual detector is a large array CCD camera. For shots giving high yields of neutrons, the camera would not only be blinded by the neutrons, it would be damaged. To get around this problem, an image plate (IP) is used as the detector.
机译:本文介绍了一种用于校准用于低能光谱范围的图像板灵敏度的方法。图像板,也称为光刺激发光(PSL)探测器,经常被证明是作为等离子体物理研究的检测器的有价值的工具。动态范围大,停止功率高,对中子损坏的耐受的优点有时超过了分辨率有限的问题和所需的远程处理。当X射线源生产高中源时需要中子损伤阻力。静态X射线成像器(SXI)是LLNL的国家点火设施(NIF)目标腔室的关键诊断,用于确定激光束的对称性。 SXI对于从DANTE诊断中正确解释数据以确定X射线辐射温度1。它由位于目标室的顶部和底部的两个诊断组成。通常的探测器是一个大型阵列CCD相机。对于提供高产量的中子产量的镜头,相机不仅会被中子蒙蔽,它将被损坏。为了解决这个问题,将图像板(IP)用作检测器。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号