首页> 外文会议>International Conference on 3D Radiation Dosimetry >Clinical experience with Mobius FX software for 3D dose verification for prostate VMAT plans and comparison with physical measurements
【24h】

Clinical experience with Mobius FX software for 3D dose verification for prostate VMAT plans and comparison with physical measurements

机译:Mobius FX软件的临床经验,用于3D剂量验证前列腺VMAT计划和物理测量的比较

获取原文

摘要

MobiusFX, an add-on software module from Mobius Medical Systems for IMRT and VMAT QA, uses measurements in linac treatment logs to calculate and verify the 3D dose delivered to patients. In this study, 10 volumetric-modulated arc therapy (VMAT) prostate plans were planned and delivered in a Varian TrueBeam linac. The plans consisted of beams with 6 and 10 MV energy and 2 to 3 arcs per plan. The average gamma value with criterion of 3% and 3mm MobiusFX and TPS: 99.96%, 2% and 2mm MobiusFX and TPS: 98.70 %. Further comparison with ArcCheck measurements was conducted.
机译:MobiusFX,来自Mobius Mevice Systems for IMRT和VMAT QA的加载版软件模块,使用LINAC处理日志中的测量来计算和验证为患者提供的3D剂量。在这项研究中,计划并在Varian Truebeam Linac中计划和交付10个容量调制的电弧疗法(VMAT)前列腺计划。该计划包括具有6和10 mV能量的光束,每个计划2至3个弧。平均伽马值具有3%和3mm MobiusFX和TPS的标准:99.96%,2%和2mm MobiusFX和TPS:98.70%。进行与ArcCheck测量的进一步比较。